BEST Triple-Resonance Setup Guide

Overview

Six Bruker BEST experiments cover a complete backbone assignment (HN, N, Cα, Cβ, CO) for a 13C/15N-labeled protein. BEST pulses excite only amide protons. Aliphatic protons and water stay untouched and speed up HN relaxation, so the recycle delay can be very short (0.2–0.4 s).

ExperimentPulse programPeaks at (H, N) of residue i13C carrier (O2P)Role
BEST-HSQCb_hsqcetf3gpsi(H, N) onlynot usedReference 2D, sample check
HNCOb_hncogp3d.2CO(i-1)CO, ~176 ppmMost sensitive 3D; CO link
HN(CA)COb_hncacogp3d.2CO(i), CO(i-1)CO, ~176 ppmPartner of HNCO
HNCACBb_hncacbgp3d.2Cα, Cβ of i and i-1aliphatic, ~40 ppmMain backbone experiment; residue type
HN(CO)CACBb_hncocacbgp3d.2Cα(i-1), Cβ(i-1)aliphatic, ~40 ppmPartner of HNCACB; identifies i-1
intra-HNCACBb_hncacbigp3dCα(i), Cβ(i) onlyaliphatic, ~40 ppmOptional; removes i/i-1 ambiguity

Recommended run order:

  1. BEST-HSQC (2D check of sample and HN range)
  2. HNCACB
  3. HN(CO)CACB
  4. HNCO
  5. HN(CA)CO
  6. intra-HNCACB (only if time allows or i/i-1 is ambiguous)

For a well-dispersed folded protein (for example a ubiquitin fold, up to ~150 residues), steps 1–5 are normally enough. 4D experiments are only worth the effort for helical, large, or disordered proteins with heavy overlap.

The simultaneous 13C/15N NOESY-HSQC (for NOE distance restraints) has its own section below. It is not a BEST experiment and uses different settings, especially a longer d1.

Common settings

Use the same values in all six experiments so the spectra overlay directly. The 13C carrier is the only setting that changes between experiments (see the next section).

ParameterValue (600 MHz)Notes
ZGOPTNS-DCALC_SPCalculates all band-selective 1H pulses from cnst54 and cnst55. Required in every experiment.
cnst548.3 ppmCenter of the HN band
cnst554.2–5.0 ppmWidth of the HN band. Excited range is cnst54 ± cnst55/2. Keep the upfield edge at or above ~5.8 ppm so Hα and water are not touched.
cnst21~176 ppmCO offset
cnst22~53–56 ppmCα offset (HNCO, HN(CA)CO, HN(CO)CACB). 53.2 ppm is fine.
O1P4.7 ppmWater
O3P116–118 ppmMatch your 15N HSQC
d10.2–0.4 sThe main time saving of BEST. Do not use 1 s.
ns8 (multiple of 8)ds ≥ 16. HSQC allows ns = 1 × n.
F3 SW~12 ppmCovers about −1.3 to 10.7 ppm with O1P at 4.7 ppm
F3 TD~680Gives aq ≈ 47 ms
aq≤ 50 ms15N decoupling runs during acquisition and d1 is short, so long aq heats the sample and probe.

The relation aq = TD / (2 × SW in Hz) controls acquisition time. To shorten aq, lower TD or increase SW. Reducing SW makes aq longer.

After changing cnst54, cnst55 or any carrier, open ased or start zg once so -DCALC_SP recalculates p41, p42, p43 and their powers. At 600 MHz they should be close to 2.2, 1.4 and 1.7 ms.

NS rule

  • ns must be a whole multiple of the phase cycle. Otherwise artifacts and axial peaks are not cancelled.
  • All five 3D experiments (HNCO, HN(CA)CO, HNCACB, HN(CO)CACB, intra-HNCACB): ns = 8 × n (8, 16, 24 …).
  • BEST-HSQC: ns = 1 × n (any value).
  • ds ≥ 16 in all experiments.
  • S/N grows with √ns. Doubling S/N needs 4 × ns and 4 × the time.
  • The same rule applies with NUS. NUS removes increments, not scans.

Per-experiment settings

The 13C carrier (O2P) is the most common setup error. CO experiments use ~176 ppm and Cα/Cβ experiments use ~40 ppm. Always check O2P when copying parameters from another experiment.

ExperimentO2PZGOPTNSExtra constantsCheck before running
BEST-HSQCnot used-DCALC_SP -DLABEL_CNcnst4 = J(NH), ~92 Hz-DLABEL_CN adds a 13C 180° pulse in t1. Use it for 13C/15N-labeled samples, omit it for 15N-only samples.
HNCO~176 ppm (CO)-DCALC_SPcnst22 ~53–56 ppmO2P must be on CO, not aliphatic.
HN(CA)CO~176 ppm (CO)-DCALC_SPcnst22 ~53–56 ppmp24 and sp9 (selective Cα 180°) must be non-zero with a shape assigned. Weakest experiment.
HNCACB~40 ppm (aliphatic)-DCALC_SPcnst21 ~176 ppmNever add -DLABEL_CB.
HN(CO)CACB~40 ppm (aliphatic)-DCALC_SPcnst21 ~176, cnst22 ~53–56 ppmNever add -DLABEL_CB.
intra-HNCACB~40 ppm (aliphatic)-DCALC_SPcnst21 ~176 ppmNever add -DLABEL_CB. 15N is full constant time, so td2 has a maximum.

Flag notes

  • -DLABEL_CB (HNCACB, HN(CO)CACB, intra-HNCACB) doubles d28 from 3.6 to 7.2 ms. Cα is then suppressed and only Cβ remains (plus Gly Cα). This is wrong for normal backbone assignment.
  • -DLABEL_CN only exists in BEST-HSQC. In the 3D experiments it has no effect, so leaving it in ZGOPTNS is harmless.
  • -DTRIMP does not exist in any of these six. It belongs to other sequences such as noesyhsqcgpsismsp3d.
  • Fixed delays (d22, d23, d25, d26, d28) are written into each pulse program. Do not change them.

Expected peak signs

  • HNCACB, HN(CO)CACB, intra-HNCACB: Cα and Cβ have opposite signs. Phase Cα positive. BEST-HN(CO)CACB differs from the classic CBCA(CO)NH, where Cα and Cβ usually have the same sign.
  • HN(CA)CO: CO(i) and CO(i-1) have the same sign. CO(i) is usually stronger.

After changing O2P or loading a parameter set, run getprosol, then check in ased that no shaped pulse length (p13, p14, p24, p41–p44) is 0.

Indirect dimensions and NUS

In all five 3D experiments F1 is 13C (States-TPPI) and F2 is 15N (Echo-Antiecho). Keep O3P, 15N SW and td2 identical across experiments, and keep the 13C SW and td1 identical within each carrier group, so strips overlay.

DimensionSW (600 MHz)Conventional TDTD with 20–25% NUSLimit
13C aliphatic (HNCACB group)70–75 ppm64–80~160 (t1max ~7 ms)Cα–Cβ coupling (~35 Hz) makes t1max beyond ~7–8 ms of little use
13CO (HNCO group)~12 ppm (170–182 ppm)6496–128 (t1max 26–35 ms)Cα is decoupled, so CO can be sampled long
15N (HNCO, HN(CA)CO, HNCACB, HN(CO)CACB)30–40 ppm32–4080–96Semi-constant time; in29/in30 are calculated automatically
15N (intra-HNCACB)30–40 ppm32–40≤ 108–144Full constant time. td2 max = 2 × d30 / in30 (~108 at 30 ppm, ~144 at 40 ppm). TopSpin reduces td2 automatically.

NUS guidance

  • Experiment time scales almost linearly with the NUS fraction.
  • Use NUS to buy resolution, not only time. With the same ~2 h per experiment, a 20% NUS grid can be about twice as long in both indirect dimensions as a conventional grid.
  • 20% NUS with ns = 8 is fine for HNCO, HNCACB and HN(CO)CACB.
  • For the weak experiments (HN(CA)CO, intra-HNCACB), check the test plane first. If weak peaks are close to noise, use 30% NUS or ns = 16.
  • Very small grids (for example td1 80 × td2 40) reconstruct less reliably. For those, use 50% NUS or full sampling.
  • In TopSpin set FnTYPE = non-uniform_sampling and NusAMOUNT, then generate the NUS list. TD still describes the full grid.

Check the total time with expt before starting.

Quick tests before the full 3D

Each test takes a few minutes and catches most setup errors. Fix rg after the first rga and keep rg, ns and ds identical when comparing settings.

  1. 1D (td1 = td2 = 1). Confirm HN signals across the expected range and low residual water. If peaks stop at a fixed ppm, check F3 SW/O1P and the cnst54/cnst55 band.
  2. 1H–15N plane (td1 = 1, td2 ≈ 64). Looks like a BEST-HSQC. Confirms the 15N carrier and SW.
  3. 1H–13C plane (td2 = 2, td1 ≈ 64). Confirms 13C carrier and SW and the peak pattern. For the HNCACB group, Cα and Cβ must both appear with opposite signs. For HN(CA)CO, two CO peaks per HN should appear.

Use td2 = 2 rather than 1 for the 13C plane. F2 is Echo-Antiecho, and td2 = 2 records both echo and antiecho. Process it as 3D and read out the F1–F3 plane.

A td2 = 1 plane can be read with rser2d and processed with xfb. It is pure phase at t2 = 0 but uses only half the scans.

Troubleshooting

SymptomLikely causeFix
No or very weak signal in HNCO / HN(CA)COO2P left at ~40 ppm from an aliphatic experimentSet O2P = cnst21 ≈ 176 ppm, rerun getprosol
No or very weak signal in HNCACB groupO2P left at ~176 ppm from a CO experimentSet O2P ≈ 40 ppm
Only Cβ peaks (plus Gly Cα)-DLABEL_CB in ZGOPTNSRemove -DLABEL_CB
HN(CA)CO shows one peak per HN, matching HNCOCα→CO transfer failingCheck p24 and sp9 are set and non-zero
Peaks missing at the edge of the HN rangeHN band too narrow, or F3 SW too smallWiden cnst55 or shift cnst54; check F3 SW and O1P
A flag change makes no differenceFlag misspelled (no leading dash) or the related pulse length is 0Write -DNAME exactly; check the pulse length in ased
zg error on in29 or td2SW combination gives a negative delay, or td2 above the constant-time limitCheck in29 ≥ 0; reduce td2 or 15N SW
Weak fast-exchanging amidesHigh pH or temperatureCheck sample conditions; BEST itself does not saturate water
Sample or probe heatingaq too long with short d1